Agilent 8453 (G1103A)

Espectrofotómetro de UV/Vis Agilent 8453 (G1103A). Rango longitud de onda de 190 a 1100 nm. Detector fotodiodoarray. Lectura en menos de 1 seg. Incluye PC con software integrado UV/VisChemstation B4: Avanzado, Bioquimico, Pack de Disolución y General. Manual de instrucciones. 

Más detalles

SP047

5 500,00 € sin IVA

Agilent 8453 Spectrophotometer Specifications

Wavelength range 190 - 1100 nm
Slit width 1 nm
Resolution > 1.6
Stray light

< 1.0 % At 200 nm, solution of 1.2% KCl, blank scan on air, 5 s integration

time; (EP*method)**

< 0.05 % At 220 nm, solution of 10 g/l NaI, blank scan on air, 5 s integration

time; (ASTM method)

< 0.03 % At 340 nm, solution of 50 g/l NaNO2, blank scan on air, 5 s integration

 

time; (ASTM method
Wavelength accuracy < ± 0.5 nm
Wavelength reproducibility < ± 0.02 nm

 

Photometric accuracy < ± 0.005 AU

 

Photometric noise < 0.0002 AU rms

 

Photometric stability < 0.001 AU/h
Baseline flatness < 0.001 AU rms
Typical scan time 1.5 s
Shortest scan time 0.1 s
Time until next scan 0.1 s

 

Software: UV/Vis Gemstation B04

  • G1116 Avanzado para el desarrollo de metodos de espectroscopía y métodos de validación de apoyo. 
  • G1117 Análisis bioquímico para kinetica de enzymas y desnaturalización térmica. (Temperatura de fusión de ADN y, desnaturalizacion de proteinas) 
  • G1118 Prueba de disolución aporta medidas manuales dirigidas a completar los sistemas robóticos automáticos
  • G1115 General del nucleo del sofware para tareas básicas y validación instrumental.  

No hay comentarios de clientes por ahora.