Biorad GenePulser XCell Maximize

Biorad GenePulser XCell

Biorad GenePulser XCell electroporation system. Composed of: main module, CE module, PC module and Shockpod cuvette chamber. Voltage from 10 to 3.000 V. Digital. Includes user manual.

More details

PS037

3 850,00 € tax excl.

Specifications:

Gene Pulser Xcell™ Total System
Includes main unit, CE module, PC module, and ShockPod™ cuvette chamber
 
Outputs Waveform: Exponential or square
Voltage: 10–3,000 V
Capacitance 10–500 V, 25–3,275 µF in 25 µF increments
500–3,000 V, 10, 25, 50 µF
Resistance (parallel) 50–1,000 Ω in 50 Ω increments, plus infinity
Sample resistance 20 Ω minimum at 10–2,500 V
600 Ω minimum at 2,500–3,000 V
Square-wave timing  10–500 V: 0.05–10 ms in 0.05 ms increments, 10–100 ms pulse in 1 ms increments, 1–10 pulses, 0.1–10 sec interval
500–3,000 V: 0.05–5 ms in 0.05 ms increments, 1–2 pulses, 5 sec minimum interval
Gene Pulser Xcell Eukaryotic System
Includes main unit, CE module, and ShockPod cuvette chamber
 
Outputs and other specifications Same as total system without the parallel resistance
Gene Pulser Xcell Microbial System
Includes main unit, PC module, and ShockPod cuvette chamber
 
Outputs Waveform: Exponential or square
Voltage: 200–3,000 V
Capacitance  10, 25, 50 µF
Resistance (parallel)  50–1,000 Ω in 50 Ω increments, plus infinity
Sample resistance  20 Ω minimum at 200–2,500 V
600 Ω minimum at 2,500–3,000 V
Square-wave timing  0.05–5 ms pulse duration in 0.05 ms increments,
1–2 pulses, 5 sec minimum interval
Gene Pulser Xcell Main Unit  
Outputs Waveform: Exponential or square
Voltage: 200–3,000 V
Capacitance 10, 25, 50 µF
Sample resistance 20 Ω minimum at 200–2,500 V
600 Ω minimum at 2,500–3,000 V
Square-wave timing  0.05–5 ms pulse duration in 0.05 ms increments,
1–2 pulses, 5 sec minimum interval
General  
Input voltage 100–120 VAC or 220–240 VAC, 50/60 Hz
Power Maximum 240 W (during short charging periods)
Operating environment Temperature 0–35°C, humidity 0–95% (noncondensing)
Regulatory Safety EN 61010, EMC EN 61326 Class A
Dimensions (W x D x H) Main unit: 31 x 30 x 14 cm
CE module: 31 x 28 x 9 cm
PC module: 31 x 28 x 5 cm
Weight Main unit: 6.6 kg
CE module: 3.1 kg
PC module: 1.9 kg

  • Height 14 cm
  • Width 31 cm
  • Depth 30 cm
  • Weight 6,6 kg

No customer comments for the moment.

Write your review

Biorad GenePulser XCell

Biorad GenePulser XCell

Biorad GenePulser XCell electroporation system. Composed of: main module, CE module, PC module and Shockpod cuvette chamber. Voltage from 10 to 3.000 V. Digital. Includes user manual.

Write your review

3 other products in the same category: